Sharing powerful practice Probes three different ergonomic proBes enABle you to Address A full rAnge of clinicAl And morphologicAl needs each patient is different. echosens has designed its probes to ensure efficient diagnosis in all circumstances. proBe: for children →Transducer specifically designed for being placed into narrow intercostal space → a higher ultrasound frequency, 5 mhz, enabling measurements adapted for chest perimeter from 45 to 75 cm → measurement depths are adapted between 15 to 50 mm depending on children’s morphology proBe: stAndArd proBe →The m probe is designed for the general population. it is used for adults with a thoracic perimeter of more than 75 cm → Ultrasound frequency is 3.5 mhz → measurements of liver stiffness take place between 25 and 65 mm under the skin proBe: for oVerweight pAtients →a more sensitive ultrasound sensor at the frequency of 2.5 mhz has beendesigned to enhance deeper signal penetration through tissues over a 35 to 75 mm depth → Xl probe must be used on patient with a Skin capsule distance (Scd) greater than 2.5 cm. automated probe selection will recommend the probe best suited to the patient’s morphology